Currently reading: Hanna Edge® Features and Specifications by ProSciTech Pty Ltd

Hanna Edge® Features and Specifications

Hanna Edge® measures pH, conductivity, and dissolved oxygen
Hanna Edge

edge’s groundbreaking design is the culmination of Hanna’s vision, design capabilities, integrated production and world class R&D. edge is a single meter that can measure pH, Conductivity, Dissolved Oxygen and is incredibly easy to use. We put a lot of thought into edge’s ease of use.

Screen

Large Easy to Read LCD

edge features a 5.5" LCD display that you can clearly view from over 5 meters. The large display and it’s wide 150° viewing angle provide one of the easiest to read LCD’s in the industry.

Buttons

Capacitive Touch

edge features a capacitive touch keypad that gives a distinctive, modern look. Since the keypad is part of the screen, your buttons can never get clogged with sample residue. The up and down keys move faster when continuously held (ideal for scrolling through numerous logs).

Fulltext

Clear Full Text Readout

edge features clear, full text guides displayed on the bottom of the screen. There is no need to decipher scrambled abbreviations or symbols; these helpful messages guide you through every process quickly and easily.


Cradle

edge® is packed with many cutting edge features to ensure accuracy and usability.

GroupShot

Digital Electrodes

edge electrodes are highly advanced, featuring a built-in microchip that stores calibration information that is automatically retrieved by edge once the electrode is plugged in. These digital electrodes also feature an easy to plug in 3.5 mm connector so you never have to worry about the right angle or aligning pins settings.

Microchip

Sensor Check

When used with Hanna’s electrodes equipped with a matching pin, edge constantly checks the impedance of the pH measuring electrode to notify you in real time in the event of glass breakage. During calibration, Sensor Check checks the state of the junction. The reference junction is also evaluated and reported on the display.

Input

3.5 mm Probe Input

Plugging an electrode in has never been simpler, no alignments, no broken pins, just connect the 3.5 mm plug and get started.


WallMount

edge® is incredibly versatile. It can be used in a variety of ways and in a variety of environments without taking up the space of a traditional benchtop meter.

Mount

Wall Mount

Using the wall mount cradle (included), edge can be placed on a wall, leaving zero footprint on the benchtop space. The cradle has a built in connector to power edge and charge its batteries. edge's zero footprint is designed to save you valuable benchtop space.

Cradle

Cradle and Electrode Holder

edge is equipped with a benchtop cradle with an adjustable swivel electrode holder to charge and hold the edge securely in place at the optimium viewing angle.


Sides

edge® is incredibly thin and weighs only 8.8 oz., has an 8 hour battery life, and dual USB inputs.

USB

Two USB Ports

edge includes one standard USB for exporting data with a flash drive and a second micro USB port for you to connect to a computer for file export and for charging your edge when the cradle is not available.

Drive

Data Logging

edge allows you to store up to 1000 log records of data. Data sets include readings, GLP data, date and time.

Microchip

Basic Mode

You can use edge in Basic Mode, ideally for routine measurements for a simplified screen and features.




pH Specifications
Rangestandard mode: -2.00 to 16.00 pH, -2.000 to 16.000 pH; ±1000 mV
basic mode: -2.00 to 16.00 pH; ±1000.0 mV
Resolutionstandard mode: 0.01 pH; 0.001 pH
basic mode: 0.1 mV
Accuracy (@25°C/77°F)±0.01 pH; ±0.002 pH; ±0.2 mV
Calibration Points5­­ in standard mode; 3 in basic mode
Calibration Buffersstandard mode: 1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45 and two custom buffers
basic mode: 4.01, 6.86, 7.01, 9.18, 10.01
Temperature CompensationATC (-5.0 to 100.0°C)*
Electrode Diagnosticsstandard mode: probe condition, response time and out of calibration range
Glass and reference junction diagnostics ( HI 11311 & HI 12301 only)
EC Specifications
Range
EC0.00 to 29.99 μS/cm; 30.0 to 299.9 μS/cm; 300 to 2999 μS/cm; 3.00 29.99 mS/cm; 30.0 to 200.0 mS/cm; up to 500.0 mS/cm (absolute EC)**
TDS0.00 to 14.99 mg/L (ppm); 15.0 to 149.9 mg/L (ppm); 150 to 1499 mg/L (ppm); 1.50 to 14.99 g/L; 15.0 to 100.0 g/L; up to 400.0 g/L (absolute TDS)**, with 0.80 conversion factor
Salinity0.0 to 400.0 % NaCl
0.01 to 42.00 PSU
0.0 to 80.0 g/L
Resolution
EC0.01 μS/cm; 0.1 μS/cm; 1 μS/cm
0.01 mS/cm; 0.1 mS/cm
TDS0.01 mg/L (ppm); 0.1 mg/L (ppm); 1 (ppm)
0.01 g/L; 0.1 g/L (0.8 TDS factor)
Salinity0.1 % NaCl; 0.01 PSU; .01 g/L
Accuracy (@25°C/77°F)
EC±1% of reading ±(0.5 μS or 1 digit, whichever is greater)
TDS±1% of reading ±(0.03 ppm or 1 digit, whichever is greater)
Salinity±1% of reading
Calibration
EC:1 point offset calibration (0.00 μS/cm in air)
1 point slope calibration in EC standard 84 μS/cm, 1413 μS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm and 118.8 mS/cm
TDS:through EC calibration
Salinity:1 point with HI 7037L 100% NaCl sea water standard (other scales through EC calibration)
Temperature CompensationnoTC, ATC (0.0 to 100.0°C)
TDS Factor0.40 to 0.80
DO Specifications
Range:0.00 to 45.00ppm; 0.0 to to 300.0% saturation
Resolution:0.01ppm; 0.1% saturation
Accuracy:± 1.5% of reading ±1 digit
Calibration Points:one or two points at 0% (HI 7040 solution) and 100% (in air)
Temperature Compensation:ATC (0 to 50°C)*
Salinity Compensation:0 to 40 g/L (with 1 g/L resolution)
Altitude Compensation:-500 to 4000m (with 100m resolution)
Temperature Specifications
Range:-20.0 to 120.0°C
Resolution:0.1°C
Accuracy:±0.5°C
Additional Specifications
Logging:stores up to 1000 records: 200 records (log-on-demand and stability logging); 600 records interval logging
Connectivity:1 USB port for storage; 1 micro USB port for charging and PC connectivity
Environment:0 to 50°C; RH max 95% non-condensing
Power Supply:5 VDC adapter (included)
Dimensions:202 x 140 x 12.7mm
Weight:250g

* temperature limits will be reduced to actual probe/sensor limits
** with temperature compensation function disabled


The EHI2020-02 kit is supplied with benchtop docking station with electrode holder, wall mount cradle, EHI11310 glass body refillable pH electrode with built in temperature sensor, two sachets each of pH 4, 7, 10, cleaning and storage solutions, USB cable, power adapter and instruction manual.
Replacement electrodes:
EHI11310 Glass body, double junction, refillable pH electrode with temp sensor
EHI11311 Glass body, double junction, refillable pH electrode with temp sensor and matching pin

EHI12300 PEI body, double junction, gel filled pH electrode with temp sensor
EHI12301 PEI body, double junction, gel filled pH electrode with temp sensor and matching pin
EHI10530 Glass body, conical tip, triple ceramic junction, refillable pH electrode with temp sensor for semi-solid products, emulsions, soil samples and low conductivity solutions
EHI10430 Glass body, single ceramic junction, refillable pH electrode with temp sensor for use with hydrocarbon, paints, solvents, seawater, strong acid and base, high conductivity solutions and tris buffers

The EHI2030-02 kit is supplied with benchtop docking station with electrode holder, wall mount cradle, EHI763100 conductivity probe, three sachets each of 1413μS and 12,880μS calibration solutions, USB cable, power adapter and instruction manual.
Replacement electrode: EHI763100 Four ring conductivity probe with built in temperature sensor

The EHI2040-02 kit is supplied with benchtop docking station with electrode holder, wall mount cradle, EHI764080 dissolved oxygen probe, EHI7041S refill electrolyte solution, two DO membrane caps and O-rings, USB cable, power adapter and instruction manual.
Replacement electrode:
EHI764080 Clark type polarographic dissolved oxygen probe

Was this article helpful?